來源: 發布時間:2022-01-08

氧氮氫分析儀的常見故障及解決方法:氧和氮空白值超過20。這是由于氣流小,不能將爐子中的空氣驅趕出去。可調節氣體流量,調節載氣壓力在0.2~0.4MPa。接通儀器載氣,放一個石墨坩堝在下電極上,打開主電源開關,點擊軟件上的關爐按鈕,關閉爐子并等待10s。調節流量調節器,直到流量計a顯示為30L/h,打開爐子。調節調節器直到流量計b顯示為50L/h,再次關閉爐子。如果以上設置不穩定,則增加流量至100L/h,反復調節直至儀器穩定。試樣加載過程中完全密閉,消除大氣的中可能帶來的雜質的污染。專業元素分析儀

專業元素分析儀,分析儀

SummaryThedeterminationoftheamountofoxygen,nitrogen,andhydrogeniniron,steel,nickel-,andcobalt-basealloysrepresentssomeofthemostimportantqualitymetricsforthesematerials.Oxygenisusedtocreatesteelfrompigironbyremovingexcesscarbon.Oxygencontentmustbecontrolledtolimittheamountofcarbonmonoxidethatcanbeformedduringsolidificationwhichmaycauseexcessiveporosity.Nitrogenisconsideredbothanimpurityaswellasanimportantalloyingagent.Itcanbepresentasanitrideorinterstitiallyinitsgaseousform.Increasednitrogencontentisknowntoincreaseyieldandtensilestrength,thusdecreasingductilityandformability.Excessivelevelsmayevolveduringsolidificationthusincreasingporosity.Highhydrogencontentistheprimarycauseofembrittlement,blisteringandflakingduetoitshighmobilitythroughthelatticeandprovidesnopotentialalloyingbenefits.TheONH836utilizesahigh-powerelectrodefurnacetoquicklyandefficientlyreleasethetargetgasesfromwithinthesample,whichallowsforaveryrapidsimultaneousdeterminationofoxygen,nitrogen,andhydrogen.新疆美國LECO分析儀多少錢20位或100位全自動分析系統。

專業元素分析儀,分析儀

氧氮氫分析儀的常見故障及解決方法儀器啟動時顯示。沒有水流。系統分析電流切斷,分析停止。這是水流探測器不正常,水泵不工作,管道堵塞。將儀器的右面板取下,觀察水流探測器,用手擠壓補水塑料水瓶,發現水流正常,顯示正常,證明水流探測器正常,管道暢通。啟動循環水泵,但分析儀顯示沒有水流,此時可判定水泵不正常。打開水泵轉子密封口,啟動泵發現電機正常運轉,此時關閉進水,拆下水泵,發現葉輪脫落。經了解,判定是由于外部冷卻水停水,儀器內循環水溫過高(水溫應≤70℃),致使葉輪(葉輪材料PVC)熱脹并脫軸。用粘合劑粘合葉輪后再粘于葉輪軸上,待粘合劑凝固后試車,儀器運行正常。

MethodSelectionTwomethodsaredescribedinthisapplicationnote;eithermethodcanbeusedtoanalyzeiron,steel,nickel-,andcobalt-basealloys.ThePrecisionMethodisrecommendedforgeneraluseandwillprovidethebestprecisionandaccuracythroughoutthetypicalO,N,andHconcentrationsfoundinthisgroupofmetals;approximatecycletimeis3.5minutes.TheFastTrackmethodcanbeusedwherespeedofanalysisisacriticalcomponent;forexample,whenmoltenmetalisbeingsampledandresultsarerequiredintheshortestpossibletime.Thismethodwillproducesuitableresultsformostsamples;approximatecycletimeis2.25minutes.Asnotedabove,samplingandsamplepreparationarekeyelementstoaccurateO,N,andHdeterminationaswell.Itisuptotheusertodeterminewhichmethodbestmeetstheirneeds.減少了繁瑣的額外數據處理。

專業元素分析儀,分析儀

Procedure-SolidSamplesPresstheAnalyzebuttonontheinstrumentscreenagain,theloadingheadslide-blockwillcloseandthelowerelectrodewillopen.Cleantheupperandlowerelectrodemanually,or,ifapplicable,removethecrucibleandpresstheanalyzebuttontocleanwiththeautomaticcleanerAddapproximately0.05gof***GraphitePowdertoa***GraphiteCrucible.Firmlyplacethecrucibleonthelowerelectrodetiporappropriateautoloaderposition.PresstheAnalyzebuttonontheinstrumentscreen,thelowerelectrodewillcloseandtheanalysissequencewillstartandendautomatically**的動態流量補償(DFC)保證熱導池氣體流量、壓力恒定。山西專業分析儀廠家報價

Cornerstone氧氮氫分析軟件同時支持***的數據文檔格式和可變的報告格式。專業元素分析儀

ONH836系列:氧氮氫分析儀(熔融法)力可ONH836系列儀器采用惰性熔融法,提供各種無機材料、鋼鐵、有色金屬、硬質合金、陶瓷材料中寬范圍的氧氮氫同時分析。ONH836系列儀器通過***采納客戶的反饋意見和創新設計,力可***儀器采用觸摸屏技術和在線軟件平臺,增強了儀器的可操作性、降低了分析成本、節省實驗室桌面空間。用戶友好界面的Cornerstone?軟件力可的Cornerstone氧氮氧分析軟件采用觸摸屏界面使用戶可完整的達到分析控制、方法設定、在線診斷、數據報告。Cornerstone氧氮氧分析軟件允許用戶在單一分析界面中完成日常的所有工作,快速而且簡單易用。力可創新的批次和重復樣品數據分組功能極大簡化了樣品結果輸出和結果標準偏差的自動計算,減少了繁瑣的額外數據處理。專業元素分析儀